POSITION
DETECTION OF PERMANENT MAGLEV
IMPELLER
IN CENTRIFUGAL HEART PUMP
Qian Kunxi Wan Fukai Ru Weimin
Zeng Pei Yuan Haiyu
(Institute of Biomedical Engineering,
Jiangsu University, Zhenjiang 212013)
|
|
Abstract: A novel permanent maglev impeller blood pump has been
developed recently, in which the impeller and the rotor are levitated by
passive magnetic and hydrodynamic bearings, therefore neither position
measurement nor feedback control is necessary. Nevertheless, the authors
detect the position of the impeller with Hall sensors in order to verify
whether the impeller is actually levitated and to certify which factors
affect the stability of the levitation. The results demonstrate that the
higher the rotating speed and the flow rate, the more stable the
equilibrium of the levitation; and the higher the viscosity of the
fluid, the more stable the suspension of the rotor; besides, it seems
that the gap between the rotor and the stator has an optimal value about
1% of the rotor diameter, while the maximal gap by hydrodynamic bearing
alone is only 0.1% of the rotor diameter.
Key words: Rotary machine Mechanical wear Permanent maglev
Rotor position detection
CLC No: R318.11
国家自然科学基金资助项目(39970736).
Received 20040312, received in revised form 20040820
|